Difference between revisions of "Firmware-633 Command-058"
(Copied from Firmware-73 Command-058, revision 1954) |
|||
Line 1: | Line 1: | ||
==DUMP OUTPUT (58, 3Ah)== | ==DUMP OUTPUT (58, 3Ah)== | ||
− | The DUMP OUTPUT command can be used to get the actuator output status. The command can be used in both host and stand-alone modes of operation. Upon completion of the command the digitizer sends out the same message as the [[Firmware- | + | The DUMP OUTPUT command can be used to get the actuator output status. The command can be used in both host and stand-alone modes of operation. Upon completion of the command the digitizer sends out the same message as the [[Firmware-633 Command-048 |SET OUTPUT command]] but with the command ID byte set to 58 (3Ah). |
The [BODY] of the DUMP OUTPUT command consists of a single 7-bit byte with the following format: | The [BODY] of the DUMP OUTPUT command consists of a single 7-bit byte with the following format: |
Latest revision as of 21:50, 11 January 2018
DUMP OUTPUT (58, 3Ah)
The DUMP OUTPUT command can be used to get the actuator output status. The command can be used in both host and stand-alone modes of operation. Upon completion of the command the digitizer sends out the same message as the SET OUTPUT command but with the command ID byte set to 58 (3Ah).
The [BODY] of the DUMP OUTPUT command consists of a single 7-bit byte with the following format:
0000xyyy: | x = 1; apply the command to all actuator outputs (yyy is ignored), ie. send 8 messages with the status for each actuator output |
yyy = [0..7]; actuator output number, where the first actuator output number = 0, and the last (8th) actuator output number = 7 |
Example: